Ionization Gauge Type 8138/29D20

The 8138/29D20 is a directly heated triode ionization gauge intended for the continuous measurement of vacuum in the range of 10-1 to 10-7 Torr. Short-term readings of up to 1 Torr can be monitored. Accidental exposure to Atmosphere at normal operating temperature will not damage the gauge. The use of a thoria-coated iridium filament makes this possible. In coating these filaments, great care must be exercised in controlling the cataphoritic process for the application of the thoria.

8138/29D20 Physical Data

Tubulation

7/8" Glass/ .750 Kovar

Envelope

7052/7056

Mounting Position

Any

Collector

.005" Nickel Mesh

Filament

Thoriated Iridium


Operating Data

Sensitivity for N2

5.5/Torr

Operating Pressure

1 x 10-7 to 1 x 10-1 Torr

Electron Bombardment De-gas

700 volts


Recommended Electrical Operating Parameters

Collector Potential

-20 to -50 volts

Grid Potential

120 to 250 volts

Filament Current

1.5 amps

Filament Voltage

1.5 to 3.5 volts

Filament Potential to Ground

+20 volts


Pin Configuration

Pins #1, 5, 9 Collector (C)

Pins #3, 7 Grid (G)

Pins #2, 6, 8 Filament Common (Fc)

Pin #4 Filament (F)

Return to Previous PageReturn HomeTurn the Page

About the Company | ETI Gauges: Principles of Operation |

Ionization Gauges | Special ETI Gauges | OEM and Custom Designs

Configurations of Standard Gauges | Earlier Model Gauges |

Replacement Filament Assemblies | ETI Custom Filaments | ETI Stems |

|
Contact ETI | Return Home |

©1999 ETI