Ionization Gauge Type 8138/29D20 The 8138/29D20 is a directly heated triode ionization gauge intended for the continuous measurement of vacuum in the range of 10-1 to 10-7 Torr. Short-term readings of up to 1 Torr can be monitored. Accidental exposure to Atmosphere at normal operating temperature will not damage the gauge. The use of a thoria-coated iridium filament makes this possible. In coating these filaments, great care must be exercised in controlling the cataphoritic process for the application of the thoria. 8138/29D20
Physical Data Tubulation 7/8" Glass/ .750
Kovar Envelope 7052/7056 Mounting
Position Any Collector .005" Nickel
Mesh Filament Thoriated
Iridium
Operating
Data Sensitivity for
N2 5.5/Torr Operating
Pressure 1 x
10-7
to 1 x 10-1
Torr Electron
Bombardment De-gas 700
volts
Recommended
Electrical Operating Parameters Collector
Potential -20 to -50
volts Grid
Potential 120 to 250
volts Filament
Current 1.5 amps Filament
Voltage 1.5 to 3.5
volts Filament Potential
to Ground +20
volts
Pin
Configuration Pins #1, 5,
9 Collector (C) Pins #3, 7 Grid
(G) Pins #2, 6,
8 Filament Common (Fc) Pin #4 Filament