Navigate through the pages by
clicking on the "Turn the Page"
button, bottom of the page.


 



2400 Philmont Ave.
PO Box 662
Huntingdon Valley, PA 19006 U.S.A.
Phone: (215) 938-4400
Fax: (215) 947-7464
Email:
techin@etigauges.com

Ionization Gauge Type 4336 and 8135

The ETI 4336 Series of Ionization Gauges are Bayard-Alpert gauges available either with a burn-out-resistant, iridium-coated filament, or two tungsten filaments. ETI was the first to incorporate a bi-filar design grid in its Bayard-Alpert gauges. Through this design, only one-half of the grid weight is suspended on each support, thereby making it more "sag resistant" during frequent outgassing conditions.

4336 Physical Data

Tubulation

3/4", 1" Pyrex or Nonex Glass or 3/4", 1" Kovar, flanges are available

Envelope

Nonex 7720 Glass

Mounting Position

Vertical*

Collector

.005" Tungsten

Filament

Thoria-Coated Iridium or Tungsten

High Stability Geometry

Available*

Grid

Tungsten helix configuration

*See Special Gauges


Operating Data

Sensitivity for N2

10/Torr

X-ray limit

2 x 10-10 Torr

Operating Pressure

2 x 10-10 to 1 x 10-3 Torr

Electron Bombardment De-gas

70 Watts

Resistance Heated De-gas

6.3 to 7.5 volts @ 10 amps

Bake-out Temperature

250° C


Recommended Electrical Operating Parameters

Collector Potential

0 volts

Shield Potential

n/a

Grid Potential

150 to 180 volts

Filament Current

4 to 6 amps

Filament Voltage

3 to 5 volts

Filament Potential to Ground

30 volts


Type 4336

4336 Single Iridium
Pin Configuration

Pin #1 Grid (G)

Pin #2 Filament (F)

Pin #3 Filament (F)

Pin #4 Grid (G)

Type 4336 T

4336T Dual Tungsten
Pin Configuration

Pin #1 Grid (G)

Pin #2 Filament #1 (F1)

Pin #3 Filament #1 (F1)

Pin #4 Filament #2 (F2)

Pin #5 Filament #2 (F2)

Pin #6 Grid (G)

Type 8135

8135 Dual Tungsten
Pin Configuration

Pin #1 Filament #1 (F1)

Pin #2 Grid (G)

Pin #3 Filament Common (Fc)

Pin #4 Grid (G)

Pin #5 Filament #2 (F2)

Return to Previous PageReturn HomeTurn the Page


About the Company | ETI Gauges: Principles of Operation |

Ionization Gauges | Special ETI Gauges | OEM and Custom Designs

Configurations of Standard Gauges | Earlier Model Gauges |

Replacement Filament Assemblies | ETI Custom Filaments | ETI Stems |

|
Contact ETI | Return Home |



©1999 ETI