Navigate
through the pages by |
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
Ionization Gauge Type 8136 The ETI 8136 is a broad range Bayard-Alpert gauge. The special fabricated gauge elements and burn-out resistant, thoria-coated iridium filament enable this gauge to operate up to 1 x 10-1 Torr. This gauge is likewise available in a long-life version that is position independent. 8136
Physical Data Tubulation 3/4", 1" Pyrex or
Nonex Glass or 3/4", 1" Kovar, flanges are
available Envelope Nonex,
platinum-coated interior Mounting
Position Vertical* Collector .005"
Tungsten Filament Thoria- Coated
Iridium High Stability
Geometry Available Grid Tungsten helix
configuration *See Special
Gauges
Operating
Data Sensitivity for
N2 8/Torr X-ray
limit 4 x 10-10
Torr Operating
Pressure 1 x 10-10 to 1 x
10-1 Torr Electron
Bombardment De-gas 70 Watts Resistance Heated
De-gas 6.3 to 7.5 volts @
10 amps Bake-out
Temperature 250°
C
Recommended
Electrical Operating Parameters Collector
Potential 0 volts Shield
Potential 0 volts Grid
Potential 150 to 180
volts Filament
Current 4 to 6
amps Filament
Voltage 3 to 5
volts Filament Potential
to Ground 30 volts
8136 Pin
Configuration Pin #1 Grid
(G) Pin #2 Filament
(F) Pin #3 Filament
(F) Pin #4 Grid
(G) Pin #5 Shield
(Platinum Coating) (S)
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||