Ionization Gauge Type 8142 Type 8142 is a versatile gauge constructed with a rugged bi-filar grid and a thoriated iridium filament. The interior of the tube is platinum-coated, shielding the gauge components from electrostatic and static charges which can occur in various operating conditions. This gauge contains a burn-out resistant, thoriated iridium filament. It is also available in a rugged, long-life design described in the special gauge section. 8142
Physical Data Tubulation 3/4", 1" Pyrex or
Nonex Glass or 3/4", 1" Kovar, flanges are
available Envelope Nonex,
platinum-coated interior Mounting
Position Vertical* Collector .005"
Tungsten Filament Thoriated
Iridium High Stability
Geometry Available Grid Tungsten helix
configuration *See Special
Gauges
Operating
Data Sensitivity for
N2 10/Torr X-ray
limit 2 x
10-10
Torr Operating
Pressure 2 x
10-10
to 1 x 10-3
Torr Electron
Bombardment De-gas 70 Watts Resistance Heated
De-gas 6.3 to 7.5 volts @
10 amps Bake-out
Temperature 250°
C
Recommended
Electrical Operating Parameters Collector
Potential 0 volts Shield
Potential Connected to
filament return Grid
Potential 150 to 180
volts Filament
Current 4 to 6
amps Filament
Voltage 3 to 5
volts Filament Potential
to Ground 30 volts
8142
Single Iridium Pin #1 Grid
(G) Pin #2 Filament
(F) Pin #3 Filament
*Platinum Coating connected to this pin