Nude Ionization Gauge Type 8140 The ETI 8140 Nude Ionization Gauge is a flange-mounted, Bayard-Alpert type for the range of 1 x 10-3 to 4 x 10-10 Torr. It is equipped with a bonded thoria, replaceable filament. The grid is of bi-filar construction. A .005"-diameter tungsten collector is provided for a low X-ray limit. The ETI 8140 is suited to either the I2R or the electron bombardment method of electrode de-gassing, making it compatible with either type of controller. This gauge may be baked to 450° C. 8140
Physical Data Connection 2-3/4" Conflat or
KF 40 flange Envelope None Mounting
Position Vertical Collector .005"
Tungsten Filament Replaceable
Thoria-Coated Iridium or Dual Tungsten High Stability
Geometry* Available in
Iridium Only Grid Tungsten helix
configuration
Operating
Data Sensitivity for
N2 10/Torr X-ray
limit 4 x
10-10
Torr Operating
Pressure 4 x
10-10
to 1 x 10-3
Torr Electron
Bombardment De-gas 70 Watts Resistance Heated
De-gas 6.3 to 7.5 volts @
10 amps Bake-out
Temperature 450°
C
Recommended
Electrical Operating Parameters Collector
Potential 0 volts Shield
Potential n/a Grid
Potential 150 to 180
volts Filament
Current 4 to 6
amps Filament
Voltage 3 to 5
volts Filament Potential
to Ground 30 volts
8140
Single Iridium 8150/8150T Pin #1
Grid Pin #1
Grid Pin #2
Grid Pin #2
Grid Pin #3
Collector Pin #3
Collector Pin #4
Filament Pin #4 Filament
#1 Pin #5
Filament Pin #5
Common Pin #6 Not
Used Pin #6 Filament
#2